Absorbance test instructions - baseline stability – Eppendorf AF2200 Plate Reader User Manual

Page 60

background image

Maintenance
Eppendorf

®

PlateReader AF2200

English (EN)

60

8.6.4

Absorbance test instructions - baseline stability

Conduct the following measurements to determine the baseline stability using the available filters.

Calculating the baseline stability

Calculate the standard deviation for each wavelength over 20 cycles.

Checking the accuracy of the absorbance

Parameter

Setting

Measuring mode

Absorbance

Wavelength

340 nm
405 nm
492 nm
590 nm
620 nm
700 nm

Number of flashes

25

Setting time

0 (μs)

Part of plate

A1

Kinetic cycles

20, minimum time interval

• No plate is required for the measurement – the plate compartment must be empty for this

measurement.

• No materials or reagents are needed for this test.

Contact Eppendorf Service to have the accuracy of the absorbance checked.