2 test instructions, upper fluorescence, Test instructions, upper fluorescence – Eppendorf AF2200 Plate Reader User Manual

Page 51

background image

Maintenance

Eppendorf

®

PlateReader AF2200

English (EN)

51

8.6.2

Test instructions, upper fluorescence

The following tests can be conducted on the device to check the device specifications for this measuring
procedure:

• Sensitivity

• Precision within a plate

• Precision from one measurement to the next

Sensitivity

Conduct the measurement with the following settings to determine the detection limit for fluorescein.

Pipette 200 μL of 1 nM fluorescein or the blank solution (0.01 M NaOH) into the corresponding wells
according to the plate layout below.

Parameter

Setting

Measuring mode

Upper fluorescence

Wavelength, excitation

485 (20 nm)

Wavelength, emission

535 (25 nm)

Number of flashes

25

Integration time

40 (μs)

Setting time

0 (μs)

Signal gain

Optimal

Plate type

GRE96fb

X

1 nM fluorescein

blank

0.01 M NaOH