Maintenance
Eppendorf
®
PlateReader AF2200
English (EN)
52
Material/reagents
• 1 nM fluorescein (in 0.01 M NaOH) (fluorescein sodium salt)
• 0.01 M NaOH (=blank solution) (NaOH pellets)
• 96-well plate
• 200 μL pipette with tips
Calculating the detection limit (sensitivity)
The result of the detection limit formula specifies the sensitivity in pM units.
Precision within a plate
Carry out the following measurement to determine the precision and reproducibility within a plate.
Concentration
F
Concentration of fluorophore in pM units
Mean value
F
Mean RFU value of the wells filled with fluorophore
Mean value
B
Mean RFU value of the wells filled with the blank
solution
Std. deviation
B
Standard deviation of the RFU values of the wells
filled with the blank solution
Parameter
Setting
Measuring mode
Upper fluorescence
Wavelength, excitation
485 (20 nm)
Wavelength, emission
535 (25 nm)
Number of flashes
25
Integration time
40 (μs)
Setting time
0 (μs)
Signal gain
Optimal
Plate type
GRE96fb
(
)
B
B
F
F
Stdev
*
3
*
mean
mean
ion
Concentrat
imit
DetectionL
−
=