Maintenance
Eppendorf

®

PlateReader AF2200

English (EN)

52

Material/reagents

• 1 nM fluorescein (in 0.01 M NaOH) (fluorescein sodium salt)

• 0.01 M NaOH (=blank solution) (NaOH pellets)

• 96-well plate

• 200 μL pipette with tips

Calculating the detection limit (sensitivity)

The result of the detection limit formula specifies the sensitivity in pM units.

Precision within a plate

Carry out the following measurement to determine the precision and reproducibility within a plate.

Concentration

F

Concentration of fluorophore in pM units

Mean value

F

Mean RFU value of the wells filled with fluorophore

Mean value

B

Mean RFU value of the wells filled with the blank
solution

Std. deviation

B

Standard deviation of the RFU values of the wells
filled with the blank solution

Parameter

Setting

Measuring mode

Upper fluorescence

Wavelength, excitation

485 (20 nm)

Wavelength, emission

535 (25 nm)

Number of flashes

25

Integration time

40 (μs)

Setting time

0 (μs)

Signal gain

Optimal

Plate type

GRE96fb

(

)

B

B

F

F

Stdev

*

3

*

mean

mean

ion

Concentrat

imit

DetectionL

=