beautypg.com

Eppendorf AF2200 Plate Reader User Manual

Page 56

background image

Maintenance
Eppendorf

®

PlateReader AF2200

English (EN)

56

Pipette 200 μL of 25 nM fluorescein or the blank solution (0.01 M NaOH) into the corresponding wells
according to the plate layout below.

Material/reagents

• 25 nM fluorescein (in 0.01 M NaOH) (fluorescein sodium salt)

• 0.01 M NaOH (=blank solution) (NaOH pellets)

• 96-well plate

• 200 μL pipette with tips

Calculating the detection limit (sensitivity)

The result of the detection limit formula specifies the sensitivity in pM units.

X

25 nM fluorescein

blank

0.01 M NaOH

Concentration

F

Concentration of fluorophore in pM units

Mean value

F

Mean RFU value of the wells filled with fluorophore

Mean value

B

Mean RFU value of the wells filled with the blank
solution

Std. deviation

B

Standard deviation of the RFU values of the wells
filled with the blank solution

(

)

B

B

F

F

Stdev

*

3

*

mean

mean

ion

Concentrat

imit

DetectionL

=