Error

Possible cause

Remedy

Measuring results are
imprecise.

• Reagent has exceeded its

shelf life.

 Ensure that the reagent is still within its

shelf life and properly prepared.

• Reagent not prepared

correctly.

 Use clean demineralized water of adequate

quality for preparation, if required.

• Pipetting incorrect.

 Ensure that the pipette is calibrated and

that pipetting is being performed correctly.

• Incubation procedure

before measurement is
incorrect.

 If the method procedure requires

incubation before the measurement,
ensure that the temperature and time for
incubation are correctly observed.

• The microplate is not

completely filled with
measuring solution, and
contains bubbles.

 Ensure that the minimum microplate

volume required for a measurement is
reached and that there are no bubbles are
in the measuring solution.

• Turbidity of the

measuring solution.

 Centrifuge turbid measuring solutions

which contain particles and use the clear
supernatant.

• PlateReader is drifting.

 Observe the ambient conditions.
 Prevent temperature changes.
 Contact Eppendorf Service.

• Plate compartment is

contaminated.

 Clean the plate compartment.

• Interfering substances

reinforce or weaken the
fluorescence signal.

 Remove the interfering substances.

• The plate compartment

flap is not closed.

 If the plate compartment flap is defective,

please contact Eppendorf Service.

Measuring results are
incorrect.

• Method incorrectly

programmed.

 Ensure that the method parameters are

entered correctly.

• Standard solution not

prepared correctly.

 Ensure that the correct standard is used

and that the measuring solution for the
standard is prepared correctly.

Positioning cannot be
completed correctly.

• The plate compartment is

heavily contaminated and
the spring mechanism is
not functioning correctly.

 Clean the plate compartment.
 Contact Eppendorf Service.