7 troubleshooting, 1 general errors, Troubleshooting 7.1 – Eppendorf AF2200 Plate Reader User Manual
Page 45: General errors, 7troubleshooting 7.1 general errors

Troubleshooting
Eppendorf
®
PlateReader AF2200
English (EN)
45
7
Troubleshooting
7.1
General errors
Error
Possible cause
Remedy
Measuring results are
imprecise.
• Reagent has exceeded its
shelf life.
Ensure that the reagent is still within its
shelf life and properly prepared.
• Reagent not prepared
correctly.
Use clean demineralized water of adequate
quality for preparation, if required.
• Pipetting incorrect.
Ensure that the pipette is calibrated and
that pipetting is being performed correctly.
• Incubation procedure
before measurement is
incorrect.
If the method procedure requires
incubation before the measurement,
ensure that the temperature and time for
incubation are correctly observed.
• The microplate is not
completely filled with
measuring solution, and
contains bubbles.
Ensure that the minimum microplate
volume required for a measurement is
reached and that there are no bubbles are
in the measuring solution.
• Turbidity of the
measuring solution.
Centrifuge turbid measuring solutions
which contain particles and use the clear
supernatant.
• PlateReader is drifting.
Observe the ambient conditions.
Prevent temperature changes.
Contact Eppendorf Service.
• Plate compartment is
contaminated.
Clean the plate compartment.
• Interfering substances
reinforce or weaken the
fluorescence signal.
Remove the interfering substances.
• The plate compartment
flap is not closed.
If the plate compartment flap is defective,
please contact Eppendorf Service.
Measuring results are
incorrect.
• Method incorrectly
programmed.
Ensure that the method parameters are
entered correctly.
• Standard solution not
prepared correctly.
Ensure that the correct standard is used
and that the measuring solution for the
standard is prepared correctly.
Positioning cannot be
completed correctly.
• The plate compartment is
heavily contaminated and
the spring mechanism is
not functioning correctly.
Clean the plate compartment.
Contact Eppendorf Service.