Maintenance
Eppendorf

®

PlateReader AF2200

English (EN)

54

Precision from one measurement to the next

Carry out the following measurement to determine the precision and reproducibility.

Pipette 200 μL of 1 nM fluorescein or the blank solution (0.01 M NaOH) into the corresponding wells
according to the plate layout below.

Material/reagents

• 1 nM fluorescein (in 0.01 M NaOH) (fluorescein sodium salt)

• 0.01 M NaOH (=blank solution) (NaOH pellets)

• 96-well plate

• 200 μL pipette with tips

Parameter

Setting

Measuring mode

Upper fluorescence

Wavelength, excitation

485 (20 nm)

Wavelength, emission

535 (25 nm)

Number of flashes

25

Integration time

40 (μs)

Setting time

0 (μs)

Signal gain

Optimal

Plate type

GRE96fb

Part of plate

A1

Kinetic

20 cycles

Time interval

Minimum

X

1 nM fluorescein

blank

0.01 M NaOH