List of serial commands, Table 17, Measurement commands – Vaisala DMT340 User Manual
Page 98: Table 18, Formatting commands, On list of serial commands on pa
![background image](/manuals/408498/98/background.png)
User's Guide _______________________________________________________________________
96 ___________________________________________________________________ M210704EN-G
List of Serial Commands
All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in bold type.
The notation
your computer keyboard. Enter a
before starting to enter commands.
In the tables below, the bold text in the brackets is the default setting.
Table 17
Measurement Commands
Command
Description
R
Start the continuous outputting
S
Stop the continuous outputting
INTV [0 ... 255 S/MIN/H]
Set the continuous output Interval (for RUN
mode)
SEND [0 ... 255]
Output the reading once
DSEND
Output the reading once from all connected
transmitters (in STOP or POLL mode)
SMODE
[STOP/RUN/POLL/MODBUS]
Set the serial interface mode
SDELAY
View or set User Port (RS232 or RS485)
answer minimum delay
SERI [baud p d s]
User Port settings (Default: 4800 E 7 1)
ADDR [0 ... 255]
Set the transmitter address (for POLL and
MODBUS modes)
NET
View or set networking parameters for LAN
and WLAN interfaces
WLAN
View or set wireless network parameters for
WLAN interface
OPEN [0 ... 255]
Open a temporarily connection to a POLL
mode device
CLOSE
Close the temporary connection (back to
POLL Mode)
Table 18
Formatting Commands
Command
Description
FORM
Set the output format of SEND and R
commands
FST
Add the state of Autocal, purge and sensor
warming in connection with SEND and R
commands
SCOM
Assign a new command name that works like
the SEND command
FDATE
Add date to R and SEND outputs
FTIME
Add time to output to SEND and R outputs