Date and time, Using display/keypad, Table 27 – Vaisala DMT340 User Manual
Page 110: Conversion factors for pressure units

User's Guide _______________________________________________________________________
108 __________________________________________________________________ M210704EN-G
where
aaaa.a = Absolute process pressure (hPa)
Example:
>pres
Pressure : 1013.00 hPa ?
>pres 2000
Pressure : 2000.00 hPa
>
Table 27
Conversion Factors for Pressure Units
From
To: hPa
mbar
1
PaN/m2
0.01
mmHg torr
1.333224
inHg
33.86388
mmH
2
O
0.09806650
inH
2
O
2.490889
atm
1013.25
at
980.665
bar
1000
psia
1)
68.94757
1) Psia = psi absolute.
Example:
29.9213 inHg = 29.9213 x 33.86388 hPa = 1013.25 hPa
Date and Time
Using Display/Keypad
If the optional Data Logger Module is installed, you can change the time
and date using the display/keypad.
1.
Press any of the arrow buttons to open the Main Menu.
2.
Select System
and press the ►arrow button to confirm your
selection.
3.
Select Date and time
and press the ►arrow button.
4.
Press the SET button to enter the adjustment mode, and use the
arrow buttons to select and change the values.
5.
You can also change the date and time formats that are shown in
the graphs. The selected formats are only used in graphical display,
they do not change the formats that are used in the serial
communication.
6.
Press EXIT to return to the basic display.