beautypg.com

Date and time, Using display/keypad, Table 27 – Vaisala DMT340 User Manual

Page 110: Conversion factors for pressure units

background image

User's Guide _______________________________________________________________________

108 __________________________________________________________________ M210704EN-G

where

aaaa.a = Absolute process pressure (hPa)

Example:

>pres
Pressure : 1013.00 hPa ?
>pres 2000
Pressure : 2000.00 hPa
>

Table 27

Conversion Factors for Pressure Units

From

To: hPa

mbar

1

PaN/m2

0.01

mmHg torr

1.333224

inHg

33.86388

mmH

2

O

0.09806650

inH

2

O

2.490889

atm

1013.25

at

980.665

bar

1000

psia

1)

68.94757

1) Psia = psi absolute.

Example:

29.9213 inHg = 29.9213 x 33.86388 hPa = 1013.25 hPa

Date and Time

Using Display/Keypad

If the optional Data Logger Module is installed, you can change the time
and date using the display/keypad.

1.

Press any of the arrow buttons to open the Main Menu.

2.

Select System

and press the ►arrow button to confirm your

selection.

3.

Select Date and time

and press the ►arrow button.

4.

Press the SET button to enter the adjustment mode, and use the
arrow buttons to select and change the values.

5.

You can also change the date and time formats that are shown in
the graphs. The selected formats are only used in graphical display,
they do not change the formats that are used in the serial
communication.

6.

Press EXIT to return to the basic display.