Chapter 4 _________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 109

Using Serial Line

To set time enter the TIME command. To set date enter the DATE
command.

TIME

DATE

These time and date settings are shown on the timestamps of PLAY
command. When you want to include time and date in the R and SEND
commands, use the FTIME and FDATE commands.

Example:

>time
Time : 13:42:49 ?

>date
Date : 2007-05-31 ?

NOTE

If the optional Data Logger Module is not installed, time and date
available on the serial ports (only) are cleared to 2000-01-01 00:00:00 at
reset or at power failure.

Data Filtering

The averaging data filter calculates a average over a certain period of
time. The lowest measurement noise is achieved with the extended
filtering. There are three filtering levels available.

Table 28

Filtering Levels

Setting

Filtering Level

OFF

No filtering.

STANDARD

Standard filtering, enabled by default.
Approximately 13 s moving average.

EXTENDED

Extended filtering. Approximately 1 min average by
default, but configurable on the serial line.


Use display/keypad to set the filtering level.

1.

Press any of the arrow buttons to open the Main Menu.

2.

Select Measuring

by pressing the ► arrow button.

3.

Select Filtering and press CHANGE to confirm your selection.

4.

Select Off/Standard/Extended and press SELECT to confirm
your selection.

5.

Press EXIT to return to the basic display.