Figure 23, Cleaning of the tightening cone – Vaisala DMT340 User Manual

Page 40

background image

User's Guide _______________________________________________________________________

38 ___________________________________________________________________ M210704EN-G

NOTE

When re-tightening the nut after detachment the nut must be tightened
without increased effort.

6.

Clean and grease the tightening cone of the fitting screw after every
tenth detachment. Change the sealing washer every time the fitting
screw is detached. Use high-vacuum grease (for example Dow
Corning) or a similar grease.

0503-033

Figure 23

Cleaning of the Tightening Cone

The numbers refer to Figure 23 above:

1

=

Fitting screw

2

=

Sealing washer

3

=

Tightening cone

4

=

Clean cotton stick

CAUTION

In pressurized processes it is essential to tighten the supporting nuts and
screws very carefully to prevent loosening of the probe by the action of
pressure.

NOTE

When the DMT340 is installed in a process with a pressure differing
from the selected operating pressure at the time of ordering, please enter
the pressure value of the process into the transmitter memory. You can
enter the value in the following ways:

- using the display/keypad option
- using the buttons on the motherboard
- using the serial line
- using the Modbus protocol

See section Pressure Compensation Setting on page 106. If you are
using Modbus, see section Configuration Registers on page 185.