Starting and configuring sensor purge, Using display/keypad (optional), Figure 73 – Vaisala DMT340 User Manual

Page 140: Sensor purge settings, Figure 74, Performing sensor purge

background image

User's Guide _______________________________________________________________________

138 __________________________________________________________________ M210704EN-G

A sensor purge should be performed always before calibration (see the
calibration instructions) or when there is a reason to believe that a sensor
has become exposed to an interfering chemical.

Starting and Configuring Sensor Purge

Using Display/Keypad (Optional)

1.

Open the Main Menu

by pressing any of the ▼▲◄► arrow

buttons.

2.

Select ►Measuring, press ►button

.

3.

Select ►Purge, press ► button.

1103-067

Figure 73

Sensor Purge Settings

- Start the sensor purge manually by selecting Start purge now

and pressing START.

- Select Purge on power-up by using the arrow buttons. Press

On/Off to turn the start-up purge on/off.

- Set the automatic purge interval by selecting Interval: ..., press

SET. Set the purge interval and the unit (minutes/hours) by
using the arrow buttons. The interval must be 10 min … 48 h.
Press OK.

- If you have the data logger module installed, you can also

specify the Date and Time when the purge is performed next.

4.

Press EXIT to return to the basic display.

1103-066

Figure 74

Performing Sensor Purge