beautypg.com

User port connection, User port operating modes, Figure 57 – Vaisala DMT340 User Manual

Page 85: Table 13, Ioioi

background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 83

User Port Connection

Use suitable serial cable between the user port RxD, GND and TxD
screw terminals and the PC serial port.

Table 13

Default Serial Communication Settings for the User
Port

Parameter

Value

Bauds

4800

Parity

Even

Data bits

7

Stop bits

1

Flow control

None

0506-033

Figure 57

Connection Example between PC Serial Port and User
Port

Connections to pins 4,6,7 and 8 are required only if your software
requires hardware handshaking.

User Port Operating Modes

When the transmitter is powered up, it will behave according to its
configured operating mode:

- In STOP mode, the transmitter outputs the software version and the

command prompt (if echo is on).

- In RUN mode a measurement output starts immediately.

- In POLL or MODBUS mode, the transmitter does not output anything

after power-up.

For a description of the modes, see section SMODE on page 117.

NOTE

RS-232 User Port cannot be used when a communication module
(LAN, WLAN, or RS-422/485 interface) has been installed.

1 2 3 4 5

6 7 8 9

+

RxD
GND
TxD
Ch1+
Ch1–
Ch2+
Ch2–

ANAL

OG

O

UT

PUT

S

US

ER P

ORT

POWE

R

RS

-232

10…

36 V

=

24 V

~

IOIOI