FILT

Use the serial line command FILT to set the filtering level.

FILT [level] [extfactor]

where

level

= Filtering level. The options are:


OFF (no filtering)
ON (standard filtering, enabled by default.
Approximately 13 s moving average)
EXT (extended filtering, approximately 1 min average by
default but configurable by changing the extfactor)

extfactor = Weighted average factor for extended filtering level.

Range 0 ... 1, default 0.030.

In extended filtering mode, the output is calculated based
on the following formula:
[(new result * extfactor) + (old result * (1.0 - extfactor))]
This means that with extfactor value 1 the transmitter
will take only the latest measurement into account, but
with extfactor value 0.1 a new output is a combination of
the previous output (90%) and the latest measurement
(10%).


Example (enable extended filtering):

>filt ext
Filter : EXT
>