beautypg.com

Operating environment, Inputs and outputs – Vaisala DMT340 User Manual

Page 168

background image

User's Guide _______________________________________________________________________

166 __________________________________________________________________ M210704EN-G

ppmv (dry)
Measurement range (typical)

10 ... 2500 ppm

Accuracy (at + 20 °C, 1 bar)

1 ppm + 20% of reading

Other measurement parameters available (depends on model)
mixing ratio, absolute humidity, pressure dewpoint calculated to 1 bar,
temperature difference (T-Td), water vapor pressure

Operating Environment

Mechanical durability

up to +180 °C (+356 °F)

for transmitter body

-40 ... +60 °C (-40 ... +140 °F)

with display

0 ... +60 °C (32 ... +140 °F)

Storage temperature range

without display

-55 … +80 °C (-67 … +176 °F)

with display

-40 … +80 °C (-40 … +176 °F)

Pressure and temperature range for probes

see probe specifications

Measured gases

non corrosive

Electromagnetic compatibility

EN61326-1: Electrical
equipment for measurement,
control, and laboratory use –
EMC requirements – for use in
industrial locations

Inputs and Outputs

Operating voltage

10 ... 35 VDC, 24 VAC

with optional power supply module

100 ... 240 VAC 50/60 Hz

Default start-up time after sensor Purge and Autocal

About 6 min

Start-up time after power-up

3 s


Power consumption at 20 °C (Uin 24VDC)

RS-232

max 25 mA

Uout 2 x 0 ... 1V / 0 ... 5V / 0 ... 10V

max 25 mA

Iout 2 x 0 ... 20 mA

max 60 mA

display and backlight

+ 20 mA

during sensor purge

+ 110 mA max

Analog outputs (2 standard, 3rd optional)

current output

0 ... 20 mA, 4 ... 20 mA

voltage output

0 ... 1 V, 0 ... 5 V, 0 ... 10 V

Accuracy of analog outputs at 20 °C

± 0.05 % full scale

Temperature dependence of the analog outputs

± 0.005 %/°C full scale


External loads

current ouputs

RL < 500 ohm

0 ... 1V output

RL > 2 kohm

0 ... 5V and 0 ... 10V outputs

RL > 10 kohm

Wire size

0.5 ... 2.5 mm

2

(AWG 20 ... 14)

stranded wires recommended


Digital outputs

RS-232

RS-422/485 (optional)

LAN (optional)

WLAN (optional)

Protocols

ASCII commands

Modbus RTU

Modbus TCP