Galvanic isolation of the power supply, Installation, Third analog output – Vaisala DMT340 User Manual
Page 58: Figure 37, Galvanic isolation module, Figure 38, Third analog output on pa
User's Guide _______________________________________________________________________
56 ___________________________________________________________________ M210704EN-G
Galvanic Isolation of the Power Supply
If galvanic isolation of the power supply line from the output signals is
needed, DMT340 can be ordered with optional galvanic isolation
module. This module prevents harmful grounding loops.
Installation
NOTE
Galvanic isolation module is not needed when using the AC power
supply module. Note also that these two modules are physically
incompatible to be installed at the same time. If both modules are
installed, there is no room to close the transmitter cover completely.
0506-030
Figure 37
Galvanic Isolation Module
Number refers to Figure 37 above:
1
= Galvanic isolation module
Third Analog Output
0503-030
Figure 38
Third Analog Output
The numbers refer to Figure 38 above:
1
=
Flat cable pins
2
=
Screw terminals for signal line
3
=
Dip switches to select the output mode and range