Galvanic Isolation of the Power Supply

If galvanic isolation of the power supply line from the output signals is
needed, DMT340 can be ordered with optional galvanic isolation
module. This module prevents harmful grounding loops.

Installation

NOTE

Galvanic isolation module is not needed when using the AC power
supply module. Note also that these two modules are physically
incompatible to be installed at the same time. If both modules are
installed, there is no room to close the transmitter cover completely.

0506-030

Figure 37

Galvanic Isolation Module

Number refers to Figure 37 above:

1

= Galvanic isolation module

Third Analog Output

0503-030

Figure 38

Third Analog Output

The numbers refer to Figure 38 above:

1

=

Flat cable pins

2

=

Screw terminals for signal line

3

=

Dip switches to select the output mode and range