beautypg.com

Using buttons on the motherboard, Using serial line, Xpres and pres – Vaisala DMT340 User Manual

Page 109: Figure 66, Pressure set buttons on the motherboard

background image

Chapter 4 _________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 107

3.

Select Pressure compensation and press the ►arrow button to
confirm you selection.

4.

Press SET and enter the pressure value by using the arrow buttons.

5.

Press OK and EXIT to return to the basic display.

Using Buttons on the Motherboard

Pressure set buttons (P

chk

and P

set

) can be used to set the process pressure.

0504-052

Figure 66

Pressure Set Buttons on the Motherboard

The numbers refer to Figure 66 above:

1

=

Pressure set button

2

=

Pressure check button

Press check button (P

chk

) and a red led will flash the current pressure

setting in bar

a

.

Press the set button (P

set

) to set the pressure value. The number of presses

equals the number of bar

a

to be set (for example, three presses = three

bar

a

). After a few seconds, the red led will confirm the setting by flashing

the new pressure value.

Using Serial Line

XPRES and PRES

Command XPRES must be used if the value is changed frequently (for
example, by an automatic system that updates the value). Its value is not
retained at reset, and when set to 0; last value set with PRES is used
instead. Use the serial line and do the following:

PRES [aaaa.a]

XPRES [aaaa.a]

where