Getting the Measurement Message from
Serial Line

Starting Continuous Outputting

Enter the R command to start the continuous output of measurements.

R

Example:

>r
Tdf=-20.6 'C H2O= 958 ppmV x= 0.6 g/kg

>r
Tdf=-20.7 'C H2O= 958 ppmV T= 23.8 'C RH= 3.3 %RH

If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars ‘*’.

You can change the format of the output with the following commands:

- Outputting interval can be changed with the INTV command.

- Output message format can be changed with the FORM command.

Stopping Continuous Outputting

Use the S command to end the RUN mode. After this command all other
commands can be used. You can also press the Esc button or reset the
transmitter to stop the outputting.

S

See command SMODE to change the default (power-up) operation
mode.