beautypg.com

Grounding the cables, Figure 16, Grounding the screen of electrical cable – Vaisala DMT340 User Manual

Page 33

background image

Chapter 3 ________________________________________________________________ Installation

VAISALA ________________________________________________________________________ 31

Grounding the Cables

Ground the screen of the electrical cable properly to achieve the best
possible EMC performance.

0605-027

Figure 16

Grounding the Screen of Electrical Cable

Refer to Figure 16 above when performing the procedure below.

1.

Cut back outer sheath to desired length.

2.

Cut back screen braiding or screen foil to dimension X.

3.

Push the domed cap nut (item 1) and the seal insert with contact
socket of the gland (item 2+3) onto the cable as shown in the
diagram

4.

Bend over the screen braiding or screen foil by about 90º
(item 4).

5.

Push the seal insert with the contact socket of the gland
(item 2+3) up to the screen braiding or screen foil.

6.

Mount the lower part (item 5) on the housing

7.

Push the seal with the contact socket of the gland (item 2+3)
into the lower part (item 5).

8.

Fasten the domed cap nut (item 1) onto the lower part (item 5).