beautypg.com

Table 32, Error messages, Table 32 error messages – Vaisala DMT340 User Manual

Page 153

background image

Chapter 6 _______________________________________________________________ Maintenance

VAISALA _______________________________________________________________________ 151

Table 32

Error Messages

Error Code Error Message

Action

E0

Humidity sensor
measurement malfunction.

Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt, water,
ice or other contaminants.

E1

Humidity sensor short
circuit

E2

Humidity sensor open
circuit

Check the integrity of the humidity probe and the
probe cable.

E3

Temperature sensor open
circuit.

E4

Temperature sensor short
circuit.

Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt water, ice
or other contaminants.

E5

Temperature
measurement malfunction

E6

Temperature sensor
current leak.

E7

Internal ADC read error

Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.

E9

Checksum error in the
internal configuration
memory

Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.

E10

Internal EEPROM read
error

E11

Internal EEPROM write
error

E12 … E13

Add-on module 1 (or 2)
connection failure

Turn off the power and check the module
connection. Turn on the power.

E14

Device internal
temperature out of range

Ensure that the operating temperature is within
the valid range.

E15

Internal RAM memory or
software checksum error

Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.

E18

Internal ADC reference
voltage out of range

E19

Internal analog output
reference voltage out of
range

E20 … E22

Configuration switches for
analog output 1/2/3 set
incorrectly

Check and re-set the switches, see page 57.

E24 … E25

Internal error in add-on
module 1 (or 2)

Disconnect the power and check the module
connection.

E26

Communication module
installed in incorrect add-
on module slot

Disconnect the power and change the
communication module to another module slot.

E28 … E29

Unknown/incompatible
module installed in add-on
module slot 1 (or 2)

Ensure that the module is compatible with the
DMT340.

E30

Internal analog voltage out
of range

Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.

E31

Internal system voltage
out of range

Check that the supply voltage is correct and the
power supply can provide enough power for the
instrument.