beautypg.com

Formatting serial line message, Ftime and fdate – Vaisala DMT340 User Manual

Page 104

background image

User's Guide _______________________________________________________________________

102 __________________________________________________________________ M210704EN-G

Formatting Serial Line Message

NOTE

Instead of using the FTIME, FDATE and FST commands described in
this section, you can use the FORM command with modifiers TIME,
DATE, and STAT. See section FORM on page 104.

FTIME and FDATE

FTIME and FDATE commands will enable/disable output of time and
date to the serial line. To add time to R and SEND outputs enter:

FTIME [x]

To add date to R and SEND outputs enter:

FDATE [x]

where

x

=

ON or OFF


Example:

>send
Tdf=-20.6 'C H2O= 959 ppmV T= 23.9 'C RH= 3.3 %RH
>ftime on
Form. time : ON
>send
23:08:27 Tdf=-20.6 'C H2O= 959 ppmV T= 23.9 'C RH= 3.3
%RH
>fdate on
Form. date : ON
>send
2000-01-31 23:08:46 Tdf=-20.6 'C H2O= 960 ppmV T= 23.9 'C
RH= 3.3 %RH
>

FST

Use the FST command to include the status of purge, sensor warming
and AutoCal in output from the SEND and R commands.

FST [x]

where

x

= ON or OFF (default)