Page 186: Table 38 measurement data registers

background image

User's Guide _______________________________________________________________________

184 __________________________________________________________________ M210704EN-G

use the floating point registers to write values outside this 16-bit signed
integer range. Maximum writable value is 32767 also for configuration
registers that do not accept negative values.

Measurement Data (Read-Only)

Table 38

Measurement Data Registers

Name

Float

Integer

Unit

RH

0001…0002

0257 (×0.01)

%

T

0003…0004

0258 (×0.01)

°C

T

d

0007…0008

0260 (×0.01)

°C

T

d/f

0009…0010

0261 (×0.01)

°C

T

d/f

(atm)

0011…0012

0262 (×0.01)

°C

T

d

(atm)

0013…0014

0263 (×0.01)

°C

a

0015…0016

0264 (×0.01)

g/m³

x

0017…0018

0265 (×0.01)

g/kg

T

w

0019…0020

0266 (×0.01)

°C

H

2

O

0021…0022

0267 (×1)

ppm

V

p

w

0023…0024

0268 (×0.1)

hPa

p

ws

0025…0026

0269 (×0.1)

hPa

h

0027…0028

0270 (×0.01)

kJ/kg

ΔT

0031…0032

0272 (×0.01)

°C

a

NTP

0033…0034

0273 (×0.01)

g/m³

H

2

O

0065…0066

0289 (×1)

ppm

W

Available measurements depend on the instrument configuration. Values
may be unavailable also in case of device failure. Read status registers or
exception status outputs to check for failures.

NOTE

Only real-time data can be read from the instrument. Data that is
recorded in the transmitter’s memory (internal memory or data logger
module) cannot be read through Modbus.