beautypg.com

FUTEK VCal Sensor Verification User Manual

Page 60

background image

Futek Advanced Sensor Technology – Futek VCal™Documentation

Rev. 1.8.0

07/26/05

©Futek Advanced Sensor Technology 2005

60

.

Error

The algebraic difference between the indicated and true value of the load being measured. Error is numerically equal to correction but

opposite in sign.

Excitation

The voltage or current applied to the input terminals of the transducer.

Hysteresis

The maximum difference between the transducer output readings for the same applied load;

one reading obtained by increasing the load from zero and the other by decreasing the load from Rated Output.
Usually measured at half Rated Output and expressed as a percent of Rated Output. Measurements should be
taken as rapidly as possible to minimize Creep.

IEEE Standard 1451.4

(Institute of Electronic and Electrical Engineers) Standard for Plug and Play for Sensors and Signal Conditioning

(

http://www.ieee.org

)

.

ISO

International Organization for Standardization. (

http://www.iso.ch/iso/en/ISOOnline.frontpage

)

Linearity

The closeness of a calibration curve to a specified straight line. Linearity is expressed as the maximum deviation of any calibration point

on a specified straight line during any one-calibration cycle.

.

Load Cell

A force transducer typically employing strain gages to measure elastic deformation, which produces an output signal that is proportional

to the applied load or force.

Load

The weight, torque, pressure or force applied to the transducer.

.

Natural Frequency

The frequency of free oscillation under no-load conditions.

.

NIST

National Institute of Standards and Technology (

http://www.nist.gov

)

Non-linearity

The maximum deviation of the Calibration Curve from a straight line drawn between the No-load and Rated Load outputs, expressed

as a percentage of the Rated Output and measured on increasing load only.

Non-repeatability

The maximum difference between transducer output readings for repeated loadings under

identical loading and environmental conditions.

.

Operating Temperature

The extremes of temperature within which the transducer will operate without

permanent adverse change to any of its performance characteristics.

.

Primary Axis

The axis along which the transducer is designed to be loaded; normally its geometric centerline.

.

Rated Load (Rated Capacity)

The maximum Axial Load that the transducer is designed to measure within its specifications.

.

Rated Output

The signal (voltage) produced by the transducer. Where the output is directly proportional to excitation, the signal is expressed in

terms of (mV/V) of excitation.

.

Reference Sensor

Sensor, generally having a known, output versus load quality, at a given location or in a given organization, from which

measurements made there are derived.

The reference sensor itself must be periodically claibrated.

Repeatability (of results of measurement)

Closeness of the agreement between the results of successive measurements of the force carried out

under the same conditions of measurement.

Conditions include same procedure, observer, instrument, conditions, and location; and carried

out over a short period of time.

Repeatability Report

For the purposes of this Glossary; a document which reports the results of comparing three separate, but identical,

test for closeness of result agreement

.