beautypg.com

Vii. futek vcal™glossary of terminology – FUTEK VCal Sensor Verification User Manual

Page 59

background image

Futek Advanced Sensor Technology – Futek VCal™Documentation

Rev. 1.8.0

07/26/05

©Futek Advanced Sensor Technology 2005

59

A2LA

The American Association for Laboratory Accreditation. (

http://www. a2la2.net

)

Accuracy

Stated as a limit tolerance which defines the average deviation between the actual output versus theoretical output. In practical

transducer applications, the potential errors of nonlinearity, hysteresis, nonrepeatability and temperature effects do not normally occur
simultaneously, nor are they necessarily additive. Therefore, accuracy is calculated based upon the RMS value of potential errors, assuming a
temperature band of ± 10° F, full rated load applied, and proper set up and calibration. Potential errors of the readout, crosstalk, or creep
effects are not included.

ANSI

The American National Standards Institute. (

http://www.ansi.org

)

ASTM

The American Society for Testing and Materials. (

http://www.astm.org

)

Audit Trail

The chronological set of records that provides evidence of system activity. These records can be used to reconstruct, review and

examine transactions from inception to output of final results. The records can also be used to track the calibration history of a reference sensor
back to it’s original calibration (generally with NIST calibrated dead weights).

Average to Best Fit Output

The average of the standard deviations between the individual measured outputs and the Best-Fit Outputs for each

load point.

Axial Load

A load applied along a line concentric with the Primary Axis.

Best Fit Equation

The equation of a straight line, calculated from a set of measurement results, which attempts to minimize the differences

(usually called

residuals) between the line and the measurement results. There is more than one statistical method used, each of which may place

the straight line in a slightly different position with respect to the measurement data.

Calibration Certificate

For the purposes of this Glossary; A document which lists all of the information relevant to a Calibration (Customer info,

Sensor info, Test Conditions, Test Results, etc.), and certifies that the information is accurate within a certain limit of uncertainty.

Calibration Curve

A record of the comparison of the transducer outputs versus standard test loads.

Calibration Interval

A period of time or amount of use between calibrations determined to be adequate to ensure equipment remains reliable,

where reliable is defined as within sufficient tolerance (or uncertainty) for the tests being done.

Calibration Load

For the purposes of this Glossary; The maximum load which is to be applied during a particular Calibration.

Calibration

The comparison of transducer outputs against standard test loads.

.

Compensated Temperature

The range of temperature over which the transducer is compensated to maintain Rated Output and Zero Balance

within specified limits.

.

Correction Factor

Numerical factor by which the uncorrected result of a measurement is multiplied to compensate for systematic error.

Creep

The change in transducer output occurring with time, while under load, and with all environmental conditions and other variables remaining

constant. Usually measured with Rated Load applied and expressed as a percent of Rated Output over a specific period of time.

Crosstalk

With one component loaded to capacity, and the other unloaded, the output of the unloaded component will not exceed the percentage

specified of its full-scale capacity.

.

Deflection

The change in length along the Primary Axis of the load cell between no-load and Rated Load conditions.

Drift

A random change in Output under constant Load conditions.

VII. Futek VCal™Glossary of Terminology