FUTEK VCal Sensor Verification User Manual
Page 33
![background image](/manuals/372314/33/background.png)
Futek Advanced Sensor Technology – Futek VCal™Documentation
Rev. 1.8.0
07/26/05
©Futek Advanced Sensor Technology 2005
33
The final tab under the
Information Tabs
section of the Calibration / Verification interface is the
Test
tab. Here is where you obtain and
record your Offset readings (First and Final Unfixtured Zero Readings), and also where you will view the Uncertainty and Zero Return
values generated by your test results.
Under the Offsets section of the
Test
tab window is
where you will take your unfixtured zero readings,
both the First and the Final Zero Readings. To do
this you can either click into the text box next to the
Zero reading you wish to take and press Enter; or
you can click into the box and click on the
Un-Fixtured
button to the right of the text box. Either way your
unfixtured zero reading will appear in the text box.
Under the Results section of the
Test
tab window is
where you will come to view the Uncertainty factor
calculated for your particular test results. This is
also where you will come to view the Zero Return
value as a percentage of Full Scale.
The
Test Grid
makes up the other half of the Calibration
/ Verification interface window.
Here is where you will define which
Channel
you have
connected your Reference Sensor to, and which you
have connected your Test Sensor to.
It is also where you will define which
Units
you want to
use to measure your Load and your Output.
This is the
Test Grid
where you perform your tests and
view your outputs.
There are also buttons available here for you to
Tare
your display, take a
Gross
reading, and
Cancel
the
Current Test
.