Technical data – KROHNE IFC 050 Converter EN User Manual

Page 73

background image

TECHNICAL DATA

8

73

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

Operating conditions

Temperature

Temperature

Temperature

Temperature

Process temperature

Refer to technical data for the measuring sensor.

Ambient temperature

Depending on the version and combination of outputs.

It is a good idea to protect the converter from external heat sources such as direct

sunlight as higher temperatures reduce the life cycle of all electronic components.

Ambient temperatures below -25°C / -13°F may affect the readability of the display.

Storage temperature

-40…+70°C / -40…+158°F

Pressure

Pressure

Pressure

Pressure

Medium

Refer to technical data for the measuring sensor.

Ambient pressure

Atmosphere

Chemical properties

Chemical properties

Chemical properties

Chemical properties

Electrical conductivity

All media except for water: ≥ 5 µS/cm

(also refer to the technical data for the measuring sensor)

Water: ≥ 20 µS/cm

State of aggregation

Conductive, liquid media

Solid content (volume)

≤ 10%

Gas content (volume)

≤ 3%

Flow rate

For detailed information, refer to chapter "Flow tables".

Other conditions

Other conditions

Other conditions

Other conditions

Protection category acc. to

IEC 529 / EN 60529

IP66/67 (acc. to NEMA 4/4X)

Installation conditions

Installation

For detailed information, refer to chapter "Installation conditions".

Inlet / outlet sections

Refer to technical data for the measuring sensor.

Dimensions and weight

For detailed information refer to chapter "Dimensions and weights".

Materials

Signal converter housing

Aluminium with a polyester topcoat

Measuring sensor

For housing materials, process connections, liners, grounding electrodes and

gaskets, refer to technical data for the measuring sensor.