OPERATION

6

65

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

Application failure

F current output A

Error, operational fault in current output.

Parameter or hardware error. No

measurement possible.

Defective, replace electronic unit.

F current output C

F software user interface

Fault revealed by CRC check of operation

software.

Replace electronic unit.

F hardware settings

The set hardware parameters do not match

the identified hardware. A dialogue appears

in the display.

Answer queries in dialogue mode, follow

directions.

Defective, replace electronic unit.

F hardware detection

Existing hardware cannot be identified.

Defective, replace electronic unit.

F RAM/ROM error IO1

A RAM or ROM error is detected during the

CRC check.

Defective, replace electronic unit.

F Fieldbus

Malfunction of the Fieldbus interface.

-

Messages on the display Description

Actions

Status: F _ _ _ _ _

Application-dependent fault, device OK,
but measured values affected.

Application test or operator action
necessary.

F application error

Application-dependent fault, but device is

okay.

Group message, when errors as described

below or other application errors occur.

F empty pipe

1 or 2 measuring electrodes are not in

contact with the medium; measured value is

set to zero. No measurement possible.

Measuring pipe not filled, function

dependent on C1.3.2. Check installation. Or

electrodes completely insulated e.g. by oil

film. Clean!

F flow exceeding limit

Measuring range exceeded, filter setting

limits measured values. No message if

empty pipe.

Limitation fct. C1.2.1, increase values.

If this limit occurs sporadically in processes with air pockets, solid contents or low

conductivity, then either the limit has to be increased or a pulse filter used so as to quell

the error messages and also reduce the measurement errors.

F field frequency too high

Field frequency is not reaching steady state,

a measured flow value is still being supplied

but may have errors. Measured values are

still supplied, but they are always too low.

No message if coil broken or bridged.

If Fct. C1.1.14 settling time is on "manual",

increase the value in Fct. C1.1.15. If it is set

to "standard", set the field frequency in Fct.

C1.1.13 according to the signal converter

nameplate.

F DC offset

ADC over-ranged by DC offsets. No

measurement can be performed, the flow is

set to zero. No message if empty pipe.

For remote signal converters, check the

connection of the signal cable.

F open circuit A

Load on current output A too high, effective

current too low.

Current not correct, mA output cable has

open circuit or load too high. Check cable,

reduce load (set < 750 ohm).

F open circuit C

F over range A

The current or the corresponding measured

value is limited by a filter setting.

Check with Fct. C2.1 hardware or sticker in

terminal compartment, which output is

connected to the terminal.

If current output: extend Fct. C2.x.6 range

and Fct. C2.x.8 limitation.

If frequency output: extend values in

Fct. C2.x.5 and Fct. C2.x.7.

F over range C

F over range D

The pulse rate or the corresponding

measured value is limited by a filter setting.

Or the demanded pulse rate is too high.

F active settings

Error during the CRC check of the active

settings.

Upload backup 1 or backup 2 settings, check

and adjust if necessary.

Messages on the display Description

Actions

Status: F _ _ _ _ _

Operational fault in device, mA output ≤
3.6 mA or set fault current (depending
on the seriousness of the fault), status
output open, pulse / frequency output:
no pulses

Repair necessary.