Technical data – KROHNE IFC 050 Converter EN User Manual
Page 72
8
TECHNICAL DATA
72
IFC 050
www.krohne.com
08/2013 - 4002184002 - MA IFC 050 R02 en
Display and user interface
Display and user interface
Display and user interface
Display and user interface
Graphic display
LC display
Size: 128 x 64 pixels, corresponds to 59 x 31 mm = 2.32" x 1.22"
Ambient temperatures below -25°C / -13°F may affect the readability of the display.
Operating elements
4 push buttons for operating the signal converter when the housing is open.
4 magnet keys for operating the signal converter when the housing is closed.
Remote control
Only generic and not device-specific DDs and DTMs available!
PACTware
TM
(including Device Type Manager (DTM))
HART
®
Hand Held Communicator from Emerson Process
AMS
®
from Emerson Process
PDM
®
from Siemens
All DTMs and drivers are available free of charge from the manufacturer's website.
Display functions
Display functions
Display functions
Display functions
Operating menu
Setting the parameters using 2 measuring pages, 1 status page, 1 graphic page
(measured values and graphics are freely adjustable)
Language display texts (as
language package)
Standard: English, French, German, Dutch, Portuguese, Swedish, Spanish, Italian
Eastern Europe: English, Slovenian, Czech, Hungarian
Northern Europe: English, Danish, Polish, Finnish
Southern Europe: English, Turkish
China: English, German, Chinese
Russia: English, German, Russian
Units
Metric, British and US units selectable as required from lists for volume / mass flow
and counting, flow velocity, electrical conductivity, temperature
Measuring accuracy
Reference conditions
Medium: water
Temperature: +20°C / +68°F
Pressure: 1 bar / 14.5 psi
Inlet section: ≥ 5 DN
Maximum measuring error
±
0.5% of the measured value above 0.5 m/s; depending on the measuring sensor
±
2.5 mm/s below 0.5 m/s; independent of the measuring sensor
For detailed information and accuracy curves, refer to chapter "Accuracy".
Current output electronics: ±10 µA; ±100 ppm/°C (typically: ±30 ppm/°C)
Repeatability
±
0.1%