6

OPERATION

52

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

6.4.2 Menu B, test

No.

Function

Setting / Description

B1 simulation

B1.1

volume flow

mass flow

Depending on the selection, simulation of the volume or mass flow

according to the following sequence:

Select:

set value /

break (exit function without simulation)

Query: start simulation?

Select:

no (exit function without simulation) / yes (start simulation)

B1.2

current output A

simulation X

[X stands for one of the connection terminals A or D]

Sequence and settings similar to B1.2, see above!

For pulse output a set number of pulses are output in 1 s!

B1.3

pulse output D

B1.3

frequency out X

B1.3

limit switch D

B1.3

status output D

B2 actual values

B2.1

operating hours

Display the actual operating hours; exit the displayed function with key ^.

B2.2

act. flow speed

Display the actual flow speed; exit the displayed function with key ^.

B2.3

electr. temperature

Display the actual electronics temperature; exit the displayed function with

key ^.

B2.4

act. coil resistance

Display the actual resistance of the field coils.

B3 information

B3.1

C number

This number clearly identifies the electronics.

B3.2

Electronic Revision ER

Reference identification number, electronic revision and production date of

the device; includes all hardware and software changes

B3.3

device serial no.

Serial no. of the system.

B3.4

electronic serial no.

Serial no. of the electronics