beautypg.com

Operation – KROHNE IFC 050 Converter EN User Manual

Page 51

background image

OPERATION

6

51

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

A6 digital outputs

A6.1

max. pulse rate

Setting the maximum pulse rate.

Limit is 120% of this pulse rate.

A6.2

pulse value unit

Selection of the unit from a list, depending on the measurement

A6.3

value p. pulse

Setting for pulse output D (volume or mass value per pulse)

Setting: xxx.xxx in l or kg or the unit selected in A6.2

A7 process input

A7.1

device serial no.

Serial no. of the system (Fct. C5.1.3)

The following process input parameters are only available, if the quick access has been activated in the menu "setup /

device / quick setup" (Fct. C6.8.3).

A7.2

zero calibration

Display of actual zero calibration value.

Query: calibrate zero?

Settings see Fct. C1.1.1.

A7.3

size

Select from size table.

A7.4

GKL

Set value acc. to nameplate; range: 0.5…20

A7.5

field frequency

Setting as on measuring sensor nameplate =

line frequency x value (from the following list):

1/2; 1/4; 1/6; 1/8; 1/12; 1/18; 1/36; 1/50

A7.6

flow direction

Define polarity of flow direction.

Forwards (according to the arrow on the measuring sensor) or backwards

(in the opposite direction to the arrow)