Operation – KROHNE IFC 050 Converter EN User Manual

Page 54

background image

6

OPERATION

54

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

C1.2 filter

C1.2.1

limitation

Limitation of all flow values, before smoothing by time constant, affects all

outputs

Settings: -xxx.x / +xxx.x m/s; condition: 1st value < 2nd value

Range 1st value: -100.0 m/s ≤ value ≤ -0.001 m/s

Range 2nd value: +0.001 m/s ≤ value ≤ +100 m/s

C1.2.2

flow direction

Define the polarity of the flow direction.

forwards (according to the arrow on the measuring sensor) or backwards (in

the opposite direction to the arrow)

C1.2.3

time constant

Time constant of all flow measurements.

xxx.x s; range: 0.0…100 s

C1.2.4

pulse filter

Suppresses noise due to solids, air/gas bubbles and sudden changes in pH.

Select:

off (without pulse filter) /

on (with pulse filter) /

automatic (with new pulse filter)

Pulse filter "on":

Pulse filter "on":

Pulse filter "on":

Pulse filter "on": The change from one measurement value to the next one is

limited to the value "pulse limitation" for the total time "pulse width".

This filter allows a faster signal tracking for slowly changing flow values.

Pulse filter "automatic":

Pulse filter "automatic":

Pulse filter "automatic":

Pulse filter "automatic": The raw flow values are collected into a buffer,

covering two times the "pulse width" values. This filter is called "median"

filter.

This filter allows a better suppression of pulse shaped disturbances

(particles or air bubbles in very noisy environment).

C1.2.5

pulse width

Length of interference and delays to be suppressed on sudden changes in

flow.

Only available, if pulse filter (Fct. C1.2.4) is "on" or "automatic"

xx.x s; range for "on": 0.01…10 s or for "automatic": 0.1...20 s

C1.2.6

pulse limitation

Dynamic limitation from one measured value to the next; only if pulse filter

(Fct. C1.2.4) is "on".

xx.x m/s; range: 0.01...100 m/s

C1.2.7

low flow cutoff

Sets low flow values to "0"; affects all outputs

x.xxx ± x.xxx m/s (ft/s); range: 0.0…10 m/s (0.0...32.8 ft/s)

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

C1.3 empty pipe detection

C1.3.1

empty pipe detect

Select:

off / empty pipe [F]; empty pipe [S]; empty pipe [I]

Flow indication "= 0" when pipe is empty

C1.3.2

limit empty pipe

Only available when empty pipe activated [..] in Fct. C1.3.1.

Range: 0.0…9999 μS (set about 50% of the lowest occurring conductivity in

operation. Conductivity below this value = signal as empty pipe)

C1.3.3

act. conductivity

Only available when empty pipe activated [..] in Fct. C1.3.1.

Actual conductivity is indicated. Activation takes place only after setting

mode is exited!

C1.3.4

time constant

Only available when empty pipe activated [..] in Fct. C1.3.1.

range: xxx.x s; 0.1...100 s

A damping is set for empty pipe detection.