beautypg.com

Electrical connections – KROHNE IFC 050 Converter EN User Manual

Page 40

background image

4

ELECTRICAL CONNECTIONS

40

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

Pulse/frequency output active

• U

int, nom

≤ 20 V

• R

V

= 1 kΩ

• C = 1000 µF
• High current mechanical counter

High current mechanical counter

High current mechanical counter

High current mechanical counter
f

max

≤ 1 Hz

• Low current mechanical counter

Low current mechanical counter

Low current mechanical counter

Low current mechanical counter
I ≤ 20 mA
R

L

≤ 10 kΩ for f ≤ 1 kHz

R

L

≤ 1 kΩ for f ≤ 10 kHz

closed:
U

0

≥ 12.5 V at I = 10 mA

open:
I ≤ 0,05 mA at U

nom

= 20 V

INFORMATION!

The active output is galvanically coupled with the current output.

This output is intended to drive directly mechanical or electronic counters.

Figure 4-16: Pulse/frequency output active P