ELECTRICAL CONNECTIONS
36
IFC 050
www.krohne.com
08/2013 - 4002184002 - MA IFC 050 R02 en
4.10.2 Laying electrical cables correctly
1 For compact versions with nearly horizontally-oriented cable entries, lay the necessary elec-
tric cables with a drip loop as shown in the illustration.
2 Tighten the screw connection of the cable entry securely.
3 Seal cable entries that are not needed with a plug.
4.11 Connection diagrams of outputs
4.11.1 Important notes
• All groups are electrically isolated from each other and from all other output circuits.
• Passive mode: An external power supply is necessary to operate (activation) the subsequent
devices (U
ext
).
• Active mode: The signal converter supplies the power for operation (activation) of the
subsequent devices, observe max. operating data.
• Terminals that are not used should not have any conductive connection to other electrically
conductive parts.
Description of used abbreviations
Figure 4-12: Protect housing from dust and water
INFORMATION!
Depending on the version, the outputs must be connected passively or actively!
I
a
I
p
Current output active or passive
P
a
P
p
Pulse/frequency output active or passive
S
a
S
p
Status output/limit switch active or passive