Measurements out of specification

Simulation of the measured values

F factory settings

Error during the CRC check of the factory

settings.

-

F backup 1 settings

Error during the CRC check of the backup 1

or 2 settings.

Save active settings in backup 1 or 2.

F Backup 2 Settings

Messages on the display Description

Actions

Status: S _ _ _ _ _

Out of specification, measurement
continues, accuracy possibly less.

Maintenance required.

S uncertain measurement Device maintenance necessary; measured

values only conditionally usable.

Group message, when errors as described

below or other influences occur.

S empty pipe

1 or 2 measuring electrodes are not in

contact with the medium; measured value is

set to zero. No measurement possible.

Measuring pipe not filled, function

dependent on Fct. C1.3.2. Check

installation. Or electrodes completely

insulated e.g. by oil film. Clean!

S field coil broken

Field coil resistance too high.

Check field coil connections to the

electronic module (for remote versions:

field current cable) for open circuit / short

circuit

S field coil bridged

Field coil resistance too low.

S electronic temperature

Upper limit for the permissible electronic

temperature has been exceeded.

Ambient temperature too high, direct solar

radiation or, for C version, process

temperature too high.

S overflow counter 1

This is counter 1 or FB2 (with Profibus).

Counter has overrun and started again at

zero.

-

S overflow counter 2

This is counter 2 or FB3 (with Profibus).

Counter has overrun and started again at

zero.

-

S backplane invalid

The data record on the backplane is invalid.

The CRC check has revealed a fault.

No data can be loaded from the backplane

when replacing electronics. Save the data to

the backplane again (Service).

S field frequency too high

The field frequency is set so high that the

field current cannot settle. The measured

values displayed are too low.

Set field frequencey lower, see Fct. C1.1.13.

Messages on the display Description

Actions

Status: C _ _ _ _ _

Output values partially simulated or
fixed

Maintenance required.

C checks in progress

Test mode of the device. Measured values

are possibly simulated values or values with

fixed settings.

Message depending on the situation via
HART

®

or FDT.

C test sensor

Test function of the measuring sensor

electronics is active.

-

C simulation fieldbus

Values on the Foundation Fieldbus interface

are simulated.

-

Messages on the display Description

Actions

Status: F _ _ _ _ _

Application-dependent fault, device OK,
but measured values affected.

Application test or operator action
necessary.