ELECTRICAL CONNECTIONS

20

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

4.4 Preparing the signal and field current cables

4.4.1 Signal cable A (type DS 300), construction

• Signal cable A is a double-shielded cable for signal transmission between the measuring

sensor and signal converter.

• Bending radius: ≥ 50 mm / 2"

INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.

Figure 4-1: Construction of signal cable A

1 Stranded drain wire (1) for the inner shield (10), 1.0 mm

2

Cu / AWG 17 (not insulated, bare)

2 Insulated wire (2), 0.5 mm

2

Cu / AWG 20

3 Insulated wire (3), 0.5 mm

2

Cu / AWG 20

4 Outer sheath
5 Insulation layers
6 Stranded drain wire (6) for the outer shield (60)