OPERATION

6

67

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

Information

Messages on the display Description

Actions

Status: I _ _ _ _ _

Information (current measurement OK)

I counter 1 stopped

This is counter 1 or FB2 (with Profibus). The

counter has stopped.

If counter to continue counting, activate

"yes" in Fct. C2.y.9 (start counter).

I counter 2 stopped

This is counter 2 or FB3 (with Profibus). The

counter has stopped.

I power fail

The device was not in operation for an

unknown period of time, because the power

was switched off. This message is for

information only.

Temporary power failure. Counters did not

run during it.

I over range display 1

1st line on page 1 (2) of display limited by

filter setting.

Select in menu "display" Fct. C4.3 and/or

C4.4 the 1st or 2nd meas. page and increase

values in functions C4.z.3 range and/or

C4.z.4 limitation.

I over range display 2

I over range conductivity

The limits for the conductivity measurement

have been exceeded (>10000 µS/cm) or are

lower than permitted (<0.1 µS/cm).

If the measuring sensor has been properly

connected and filled with medium, this does

not affect the flow measurement. The

conductivity measured values cannot be

used.

I backplane sensor

The data on the backplane are not usable

because they have been generated with an

incompatible version.

-

I backplane settings

The global settings on the backplane are not

usable because they have been generated

with an incompatible version.

-

I backplane difference

The data on the backplane differ from the

data in the display. If the data are usable, a

dialogue is indicated in the display.

-

I write cycles overfl.

The maximum number of write cycles of the

EEPROM or FRAMS on the Profibus DP PCB

has been exceeded.

-

I baudrate search

Searching for the baudrate of the Profibus

DP interface.

-

I no data exchange

There is no data exchange between the

signal converter and the Profibus.

-

I conductivity off

Conductivity measurement switched off.

Change the settings in Fct. C1.3.1.

I empty pipe

1 or 2 measuring electrodes are not in

contact with the medium; measured value is

set to zero. No measurement possible.

Measuring pipe not filled, function

dependent on Fct. C1.3.2. Check

installation. Or electrodes completely

insulated e.g. by oil film. Clean!

I diagnosis channel off

Diagnosis value switched off.

Change the settings in Fct. C1.3.17.