OPERATION

6

61

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

C6.5 graphic page

C6.5.1

select range

Select:

manual (set range in Fct. C6.5.2) /

automatic (automatic depiction based on the measured values)

Reset only after parameter change or after switching off and on.

C6.5.2

range

Set the scaling for the Y axis. Only available if "manual" is set in C6.5.1.

±

xxx ±xxx%; range: -100…+100%

(1st value = lower limit / 2nd value = upper limit),

condition: 1st value ≤ 2nd value

C6.5.3

time scale

Set the time scaling for the X axis, trend curve

xxx min; range: 0…100 min

C6.6 special function

C6.6.1

reset errors

reset errors?

Select: no / yes

C6.6.2

save settings

Save current settings.

Select:

break (exit function without saving) /

backup 1 (saves the settings in the backup 1 storage place) /

backup 2 (saves the settings in the backup 2 storage place)

Query: go on with copy? (cannot be undone)

Select:

no (exit the function without saving /

yes (copy current settings to storage backup 1 or backup 2)

C6.6.3

load settings

Load saved settings

Select:

break (exit function without loading) /

factory settings (reload factory settings) /

backup 1 (loads the settings from the backup 1 storage place) /

backup 2 (loads the settings from the backup 2 storage place) /

load sensor data (restore factory setting of the values for the measuring

sensor. Display and I/O settings are retained!)

Query: go on with copy? (cannot be undone)

Select:

no (exit the function without saving /

yes (load data from the selected storage place)

C6.6.4

password quick set

Password required to change data in the quick setup menu.

0000 (= to quick setup menu without password)

xxxx (password required); range 4 digits: 0001…9999

C6.6.5

password setup

Password required to change data in the setup menu

0000 (= to quick setup menu without password)

xxxx (password required); range 4 digits: 0001…9999

C6.7 units

C6.7.1

volume flow

m³/h; m³/min; m³/s; L/h; L/min; L/s (L = litres);

cf/h; cf/min; cf/s; gal/h; gal/min; gal/s;

IG/h; IG/min; IG/s;

free unit (set factor and text in the next two functions, sequence see below)

C6.7.2

text free unit

For text to be specified refer to

Set free units

on page 63:

C6.7.3

[m³/s]*factor

Specification of the conversion factor, based on m³/s:

xxx.xxx refer to

Set free units

on page 63