Operation – KROHNE IFC 050 Converter EN User Manual
Page 61
OPERATION
6
61
IFC 050
www.krohne.com
08/2013 - 4002184002 - MA IFC 050 R02 en
C6.5 graphic page
C6.5.1
select range
Select:
manual (set range in Fct. C6.5.2) /
automatic (automatic depiction based on the measured values)
Reset only after parameter change or after switching off and on.
C6.5.2
range
Set the scaling for the Y axis. Only available if "manual" is set in C6.5.1.
±
xxx ±xxx%; range: -100…+100%
(1st value = lower limit / 2nd value = upper limit),
condition: 1st value ≤ 2nd value
C6.5.3
time scale
Set the time scaling for the X axis, trend curve
xxx min; range: 0…100 min
C6.6 special function
C6.6.1
reset errors
reset errors?
Select: no / yes
C6.6.2
save settings
Save current settings.
Select:
break (exit function without saving) /
backup 1 (saves the settings in the backup 1 storage place) /
backup 2 (saves the settings in the backup 2 storage place)
Query: go on with copy? (cannot be undone)
Select:
no (exit the function without saving /
yes (copy current settings to storage backup 1 or backup 2)
C6.6.3
load settings
Load saved settings
Select:
break (exit function without loading) /
factory settings (reload factory settings) /
backup 1 (loads the settings from the backup 1 storage place) /
backup 2 (loads the settings from the backup 2 storage place) /
load sensor data (restore factory setting of the values for the measuring
sensor. Display and I/O settings are retained!)
Query: go on with copy? (cannot be undone)
Select:
no (exit the function without saving /
yes (load data from the selected storage place)
C6.6.4
password quick set
Password required to change data in the quick setup menu.
0000 (= to quick setup menu without password)
xxxx (password required); range 4 digits: 0001…9999
C6.6.5
password setup
Password required to change data in the setup menu
0000 (= to quick setup menu without password)
xxxx (password required); range 4 digits: 0001…9999
C6.7 units
C6.7.1
volume flow
m³/h; m³/min; m³/s; L/h; L/min; L/s (L = litres);
cf/h; cf/min; cf/s; gal/h; gal/min; gal/s;
IG/h; IG/min; IG/s;
free unit (set factor and text in the next two functions, sequence see below)
C6.7.2
text free unit
For text to be specified refer to
Set free units
on page 63:
C6.7.3
[m³/s]*factor
Specification of the conversion factor, based on m³/s:
xxx.xxx refer to
Set free units
on page 63