beautypg.com

Operation, 3 menu c, setup – KROHNE IFC 050 Converter EN User Manual

Page 53

background image

OPERATION

6

53

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

6.4.3 Menu C, setup

No.

Function

Settings / descriptions

C1 process input

C1.1 calibration

C1.1.1

zero calibration

Display of actual zero calibration value.

calibrate zero?

Select:

break (return with ^ key)

standard (factory setting) /

manual (display last value, set new value,

range: -1.00…+1 m/s) /

automatic (shows the current value as the new zero calibration value)

C1.1.2

size

Select from size table.

C1.1.3

GKL

Set value acc. to nameplate; range: 0.5…20

C1.1.4

measurement

Select:

volume flow (default setting) /

mass flow (using the fixed density, the volume flow is converted into mass

flow)

C1.1.5

range

Range for the analogue outputs (current output, frequency output and

display)

range: 0.0…100%

C1.1.6

density

For calculating the mass flow from the volume flow

range: 0.1…5 kg/l

C1.1.7

target conduct.

Reference value for on-site calibration; range: 1.000...50000 μS/cm

C1.1.8

EF electr. factor

For calculation of the conductivity based on the electrode impedance.

Query: calibrate EF?

Select:

break (return with ^ key) /

standard (with factory setting) /

manual (set desired value) /

automatic (determines EF according to the setting in Fct. C1.1.10)

C1.1.9

field frequency

Setting as on measuring sensor nameplate =

line frequency x value (from the following list):

1/2; 1/4; 1/6; 1/8; 1/12; 1/18; 1/36; 1/50

C1.1.10

select settling

Select settling (special function)

Select:

standard (fixed allocation) /

manual (manual time setting for the settling time for the field current)

C1.1.11

settling time

Only when "manual" selected in Fct. C1.1.10; range: 1.0…250 ms

C1.1.12

line frequency

Setting the line frequency.

Select: 50 Hz or 60 Hz

C1.1.13

act. coil resistance

Display of the actual resistance of the field coil.