6

OPERATION

58

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

C2.3 limit switch D

C2.3.1

threshold

Switching level, set threshold with hysteresis

C2.3.2

polarity

Set polarity, please note flow direction in C1.2.2!

Select:

both polarities (plus and minus values are displayed) /

positive polarity (display for negative values = 0) /

negative polarity (display for positive values = 0) /

absolute value (always displays positive, with both negative and positive

values)

C2.3.3

time constant

Range: 000.1…100 s

C2.3.4

invert signal

Select:

off (activated output: switch closed) /

on (activated output: switch open)

C2.3.5

information

Serial no. of the I/O board, software version no. and production date of the

circuit board

C2.3.6

simulation

For sequence, see B1.3 limit switch D

No.

Function

Settings / descriptions

C3 I/O counter

C3.1

counter 1

Set function of counter.

_ stands for 1, 2 (= counter 1, 2)

C3.2

counter 2

C3._.1

funct. of counter

Select:

sum counter (counts positive + negative values) /

+counter (counts only the positive values) /

-counter (counts only the negative values) /

off (counter is switched off)

C3._.2

preset value

If this value is reached, positive or negative, a signal is generated that can be

used for a status output at which "preset counter X" has to be set.

Preset value (max. 8 digits) x.xxxxx in selected unit, see C5.7.10 and C5.7.13

C3._.3

reset counter

reset counter?

Select:

no / yes (available if activated in C5.9.1)

C3._.4

set counter

Set counter _ to the desired value

Select:

break (exit function) /

set value (opens the editor to make the entry)

Query: set counter?

Select:

no (exit function without setting the value) /

yes (sets the counter and exits the function)

C3._.5

stop counter

Counter _ stops and holds the current value

Select:

no (exits the function without stopping the counter) /

yes (stops the counter and exits the function)

C3._.5

start counter

Start counter _ after that counter is stopped

Select:

no (exits the function without starting the counter) /

yes (starts the counter and exits the function)

C3._.7

information

Serial no. of the I/O board, software version no. and production date of the

circuit board