TECHNICAL DATA
8
81
OPTISONIC 6300 P
www.krohne.com
03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en
Status output
Status output
Status output
Status output
Isolation
The output is galvanically isolated from the other circuits.
Function and settings
Settable as automatic measuring range change, indicator for direction of
flow, overflow, error, operating point or empty pipe detection
Status and/or control: ON or OFF
Operating data
Active
U
int
= 15 VDC
I ≤ 20 mA
open:
I ≤ 0.05 mA
closed:
U
0, nom
= 15 V at I = 20 mA
Passive
U
ext
≤ 32 VDC
I ≤ 100 mA
open:
I ≤ 0.05 mA at U
ext
= 32 VDC
closed:
U
0, max
= 0.2 V at I ≤ 10 mA
U
0, max
= 2 V at I ≤ 100 mA
Current inputs
Current inputs
Current inputs
Current inputs
Isolation
The inputs are not galvanically isolated from the other circuits.
Function
Input of temperature, used for energy calculation in combination with
flow measurement
Range: -50...500°C / -58...932°F (default: 0...120°C / -32...248°F)
Operating data
Active
U
int
= 15 VDC
I ≤ 22 mA
I
max
= 26 mA (electronically limited)
U
0, min
= 9 V with I ≤ 22 mA
No HART
®
Passive
U
ext
≤ 32 VDC
I ≤ 22 mA
I
max
= 26 mA (electronically limited)
U
0, max
= 5 V with I ≤ 22 mA
No HART
®
Approvals and certificates
CE
CE
CE
CE
This device fulfills the statutory requirements of the EC directives.
The manufacturer certifies successful testing of the product by applying
the CE mark.
Electromagnetic compatibility Directive: 2004/108/EC
Harmonized standard: EN 61326-1: 2006
Low voltage directive
Directive: 2006/95/EC
Harmonized standard: EN 61010: 2001
Other approvals and standards
Other approvals and standards
Other approvals and standards
Other approvals and standards
Protection category acc. to
IEC 529 / EN 60529 /
NEMA 250/2003
Sensor: IP 67 / NEMA 6
Converter: IP 65 / NEMA 4
Trunk on wheels: IP 67 / NEMA 6
Power adapter: IP 40 / NEMA 1
Shock test sensor
IEC 60068-2-27
Vibration test sensor
IEC 60068-2-64