beautypg.com

Electrical connections, 6 i/o cable – KROHNE OPTISONIC 6300 P-6400 EN User Manual

Page 46

background image

5

ELECTRICAL CONNECTIONS

46

OPTISONIC 6300 P

www.krohne.com

03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en

5.6 I/O cable

To use the available inputs and outputs of the converter, you need an optional I/O box. The I/O
box is available in two versions (standard I/O functions and heat measurement version).

• Plug the connector 2 of the I/O box 3 in the connector 1 of the converter. Then use the

connectors 4 to setup the wanted I/O, as shown in the connection diagrams on the next
pages.

Versions of I/O box

Figure 5-5: I/O box

1 I/O connector on converter
2 Connector of I/O box
3 I/O box with screw terminals for 1 mm

2

wiring

4 Screw terminal pin 1...15

Screw terminal
pin

Standard I/O box

I/O box with heat option

1

Frequency / pulse output D

Frequency / pulse output D

2

Frequency / pulse output D-

Frequency / pulse output D-

3

Status output X

Status output X

4

Status output X-

Status output X-

5

Current output C+

Current output C+

6

Current output C

Current output C

7

Current output C-

Current output C-

8

Current input A+

Temperature sensor 1

(PT 100, 4 wire connection)

9

Current input A-

10

Current input B+

11

Current input B-

12

Not connected

Temperature sensor 2

(PT 100, 4 wire connection)

13

Not connected

14

Not connected

15

Not connected

Table 5-1: Terminals of I/O box