Installation for flow measurement – KROHNE OPTISONIC 6300 P-6400 EN User Manual
Page 34
3
INSTALLATION FOR FLOW MEASUREMENT
34
OPTISONIC 6300 P
www.krohne.com
03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en
F
factory settings error during CRC check of factory
settings
F
back up 1 (or 2)
settings
error during CRC check of back up 1
(or 2) settings
F
signal lost
path 1
signal lost at path 1
check signal cable / check for pipe
obstructions
F
signal lost
path 2
signal lost at path 2
check signal cable / check for pipe
obstructions
F
pipe/sens1
param.
unrealistic parameter settings for
pipe in combination with path 1
check parameters in menu X6
F
pipe/sens2
param.
unrealistic parameter settings for
pipe in combination with path 2
check parameters in menu X6
S
unreliable 1
unreliable measurement at pipe 1
check process conditions for gas
bubbles, solids
S
unreliable 2
unreliable measurement at pipe 2
check process conditions for gas
bubbles, solids
S
zero converter
invalid value at power up
power off - wait 5 seconds - power
on device
S
overflow
counter 1 (or 2,
3)
counter is overflowing and will start
again at zero
no action needed
S
backplane
invalid
error during CRC check of
backplane
restore data records on backplane
I
counter 1 (or 2,
3) stopped
counter has stopped
reset counter in menu C8.9.1
(or C8.9.2, C8.9.3)
I
control input A
(or B) active
information only
no action needed
I
over range
display 1 (or 2)
1
st
row on 1
st
(or 2
nd
) measurement
page is limited by parameter setting
extend upper or lower limit for
limitation in menu C8.3.4
I
backplane
sensor
incompatible data sensor on
backplane
I
backplane
settings
incompatible data on backplane
I
backplane
difference
different data on backplane and
display
I
optical
interface
optical interface is operational, local
display can not be used
I
softw sync
error
incompatible DSP and
microprocessor software