Installation for flow measurement – KROHNE OPTISONIC 6300 P-6400 EN User Manual

Page 34

background image

3

INSTALLATION FOR FLOW MEASUREMENT

34

OPTISONIC 6300 P

www.krohne.com

03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en

F

factory settings error during CRC check of factory

settings

F

back up 1 (or 2)

settings

error during CRC check of back up 1

(or 2) settings

F

signal lost

path 1

signal lost at path 1

check signal cable / check for pipe

obstructions

F

signal lost

path 2

signal lost at path 2

check signal cable / check for pipe

obstructions

F

pipe/sens1

param.

unrealistic parameter settings for

pipe in combination with path 1

check parameters in menu X6

F

pipe/sens2

param.

unrealistic parameter settings for

pipe in combination with path 2

check parameters in menu X6

S

unreliable 1

unreliable measurement at pipe 1

check process conditions for gas

bubbles, solids

S

unreliable 2

unreliable measurement at pipe 2

check process conditions for gas

bubbles, solids

S

zero converter

invalid value at power up

power off - wait 5 seconds - power

on device

S

overflow

counter 1 (or 2,

3)

counter is overflowing and will start

again at zero

no action needed

S

backplane

invalid

error during CRC check of

backplane

restore data records on backplane

I

counter 1 (or 2,

3) stopped

counter has stopped

reset counter in menu C8.9.1

(or C8.9.2, C8.9.3)

I

control input A

(or B) active

information only

no action needed

I

over range

display 1 (or 2)

1

st

row on 1

st

(or 2

nd

) measurement

page is limited by parameter setting

extend upper or lower limit for

limitation in menu C8.3.4

I

backplane

sensor

incompatible data sensor on

backplane

I

backplane

settings

incompatible data on backplane

I

backplane

difference

different data on backplane and

display

I

optical

interface

optical interface is operational, local

display can not be used

I

softw sync

error

incompatible DSP and

microprocessor software