INSTALLATION FOR FLOW MEASUREMENT

22

OPTISONIC 6300 P

www.krohne.com

03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en

The best suitable sensor is automatically selected from the available sensor types in the
converter and the appropiate number of traverses is indicated. If none of the sensors is suitable,
the converter will show "Transducer set : None".

Calibration number

Compare the calibration number with the number on the
transducer cable. If needed, select a different transducer
set to change the calibration number.

Number of traverses

1 traverse = Z mode
2 traverses = V mode
4 traverses = W mode
See figure below for an explanation of the number of
traverses.

Figure 3-10: Number of traverses

1 1 traverse (Z mode)
2 2 traverses (V mode)
3 4 traverses (W mode)

Pipe

Available sensor

Traverse mode

DN15...150

Small 2 MHz, 1 rail

DN50...250

Medium 1 MHz, 1 rail

DN200...750

Medium 1 MHz, 2 rails

DN400...1500

Medium 1 MHz, 2 rails

CAUTION!
Normally, install the sensors as advised. If the quality of the pipe walls is poor and / or in case of
scaling inside the pipes, try to decrease the amount of traverses or (if possible) use a medium
sensor instead of a small sensor.