Installation for flow measurement – KROHNE OPTISONIC 6300 P-6400 EN User Manual
Page 22
3
INSTALLATION FOR FLOW MEASUREMENT
22
OPTISONIC 6300 P
www.krohne.com
03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en
The best suitable sensor is automatically selected from the available sensor types in the
converter and the appropiate number of traverses is indicated. If none of the sensors is suitable,
the converter will show "Transducer set : None".
Calibration number
Compare the calibration number with the number on the
transducer cable. If needed, select a different transducer
set to change the calibration number.
Number of traverses
1 traverse = Z mode
2 traverses = V mode
4 traverses = W mode
See figure below for an explanation of the number of
traverses.
Figure 3-10: Number of traverses
1 1 traverse (Z mode)
2 2 traverses (V mode)
3 4 traverses (W mode)
Pipe
Available sensor
Traverse mode
DN15...150
Small 2 MHz, 1 rail
DN50...250
Medium 1 MHz, 1 rail
DN200...750
Medium 1 MHz, 2 rails
DN400...1500
Medium 1 MHz, 2 rails
CAUTION!
Normally, install the sensors as advised. If the quality of the pipe walls is poor and / or in case of
scaling inside the pipes, try to decrease the amount of traverses or (if possible) use a medium
sensor instead of a small sensor.