INSTALLATION FOR ENERGY MEASUREMENT

4

39

OPTISONIC 6300 P

www.krohne.com

03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en

If another energy set is used, teach in the required values.

4.3.2 Program the process input

• Go to menu 2.4.2.3.1 via "Measurement Setup Process input Heat Function off "

• Choose "Heating" or "Cooling" to activate energy measurement.

• Choose in "Temperature input " which sensor is is located at the supply side of the process.

2.4.5.2

Function

Function

Function

Function

Extended range A 0%

Extended range A 100%

Extended range B 0%

Extended range B 100%

Temperature range A 0%

Temperature range A 100%

Temperature range B 0%

Temperature range B 100%

Time constant A

Time constant B

4 mA trimming A

20 mA trimming A

4 mA trimming B

20 mA trimming B

On

On

On

On

3.6 mA

21 mA

3.6 mA

21 mA

0.0 °C

120.0 °C

0.0 °C

120.0 °C

1 s

1 s

4 mA

20 mA

4 mA

20 mA

13-04-2010 14:11:09 13 MB free

INFORMATION!
The "Extended range A/B 0% and 100%" are meant for an alarm function. If a value is measured
that is lower than the "Extended range A/B 0%" value or higher than the "Extended range A/B
100%" value, an event will be logged in the event logger.

2.4.2.3.1

Heating

Heating

Heating

Heating

Cooling

Off

13-04-2010 14:11:09 13 MB free

2.4.2.3

Function

Temperature input

Temperature input

Temperature input

Temperature input

Sensor location

Fluid

Heating

Terminal A at supply

Terminal A at supply

Terminal A at supply

Terminal A at supply

Supply

Water

13-04-2010 14:11:09 13 MB free

INFORMATION!
In case the function is "Heating", the temperature at the "supply" side is the highest.
In case the function is "Cooling", the temperature at the "supply" side is the lowest.