INSTALLATION FOR FLOW MEASUREMENT

28

OPTISONIC 6300 P

www.krohne.com

03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en

3.4.3 1 traverse with 2 rails (DN400...1500)

Applicable diameters with two rails:

• Measure the outer diameter of the pipe with the tape measure.
• Calculate the half of the outer diameter.
• Start at the left side of the rail and locate the opposite side using the calculated half of the

outer diameter.

• Put a mark at this position.
• Repeat this at the right side of the rail.
• Draw a line between the two markings.
• Mount the DOWN rail in such a way that the transducer is at the marked location.

Number of
traverses

Diameter range

1

DN400...1500

Figure 3-13: Mounting second rail in Z mode (1 traverse) using a reference point

1 Measure the distance between the transducer of the UP rail and the reference point.
2 Add the advised distance to determine the position of the second transducer.