INSTALLATION FOR FLOW MEASUREMENT
3
21
OPTISONIC 6300 P
www.krohne.com
03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en
An advise is given in the next menu:
CAUTION!
Inaccurate input of the outside diameter will affect the accuracy of the measured flow rate.
INFORMATION!
In case of a two pipe configuration, the converter will ask if the data entered for pipe 1 has to be
used for pipe 2 too.
Menu 1.2 and 1.3 are shown again to enter the data for the second pipe.
1.3
Fluid
VoS Fluid
Viscosity
< Previous | Next >
< Previous | Next >
< Previous | Next >
< Previous | Next >
Water
1485.0 m/s
1 mm
2
/s
13-04-2010 14:11:09 13 MB free
Fluid
Select the correct fluid from the table.
VoS Fluid
Velocity of Sound of the selected fluid. Only change it if
highly accurate values are available, for instance
temperature compensated.
Viscosity
Only change it if the viscosity is well known.
Choose next
CAUTION!
Find the calibration numbers that are noted on the labels on the cable of each transducer. Make
sure that both transducers have the same calibration number as shown by the converter.
1.7
Transducer set
Calibration number
Number of traverses
< Previous | Next >
< Previous | Next >
< Previous | Next >
< Previous | Next >
Ta
522505050
2
13-04-2010 14:11:09 13 MB free
CAUTION!
Normally, do not change the settings in this menu.
INFORMATION!
In case of a two sensor configuration, this loop will be shown twice. After installing the first
transducer the converter will produce a second advice for the second installation.
Transducer set
Select the value that is on the label on the transducer
cable.
Note: A maximum of three transducer sets can be
programmed in the converter, called Ta, Tb or Tc.