5

ELECTRICAL CONNECTIONS

50

OPTISONIC 6300 P

www.krohne.com

03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en

• If the following maximum load resistance R

L, max

is exceeded, the load resistance R

L

must be

reduced accordingly by parallel connection of R:
f ≤ 100 Hz: R

L, max

= 47 kΩ

f ≤ 1 kHz: R

L, max

= 10 kΩ

f ≤ 10 kHz: R

L, max

= 1 kΩ

• The minimum load resistance R

L, min

is calculated as follows:

R

L, min

= (U

ext

- U

0

) / I

max

Status output

• U

ext

≤ 32 VDC

• I ≤ 100 mA
• R

L, max

= 47 kΩ

R

L, min

= (U

ext

- U

0

) / I

max

• open:

I ≤ 0.05 mA at U

ext

= 32 VDC

closed:
U

0, max

= 0.2 V at I ≤ 10 mA

U

0, max

= 2 V at I ≤ 100 mA

• The output is open when the device is de-energized.

Figure 5-8: Pulse frequency output active P

a

C+

C-

D

D-

INFORMATION!

Any connection polarity.

Galvanically isolated.

Figure 5-9: Status output / limit switch passive S

p