INSTALLATION FOR FLOW MEASUREMENT

3

31

OPTISONIC 6300 P

www.krohne.com

03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en

• Proceed with the optimization loop in step 4.

3.5 Step 4: Optimization loop

• Press "".

• Make sure that the transducers are mounted at the advised distance.
• Press Next .

• Press Optimize position .

• The optimization loop will give a new advise on the transducer distance. Slide the transducers

to the new position. Repeat this loop until the advised distance does not change more than 1%
or 1 mm.

1.8

Advised sensor position

52.18 mm

52.18 mm

52.18 mm

52.18 mm

Signal quality

78%

< Previous | Next >

13-04-2010 14:11:09 13 MB free

CAUTION!
If the actual sensor position is not entered accurately, the sound velocity can not be calculated
correctly and optimization of the sensor position will fail.

1.10

Advised sensor position

Actual sensor position

66.05 mm

66.05 mm

< Previous | Next >

13-04-2010 14:11:09 13 MB free

1.12

Volume flow

Velocity of sound

Signal quality

Optimize position

Optimize position

Optimize position

Optimize position

50 m

3

/h

1481.3 m/s

69%

< Previous | Next >

13-04-2010 14:11:09 13 MB free

INFORMATION!
The optimization loop uses the measured VoS and compares it to the entered VoS.