Installation for flow measurement, 5 step 4: optimization loop – KROHNE OPTISONIC 6300 P-6400 EN User Manual
Page 31
INSTALLATION FOR FLOW MEASUREMENT
3
31
OPTISONIC 6300 P
www.krohne.com
03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en
• Proceed with the optimization loop in step 4.
3.5 Step 4: Optimization loop
• Press "".
• Make sure that the transducers are mounted at the advised distance.
• Press Next .
• Press Optimize position .
• The optimization loop will give a new advise on the transducer distance. Slide the transducers
to the new position. Repeat this loop until the advised distance does not change more than 1%
or 1 mm.
1.8
Advised sensor position
52.18 mm
52.18 mm
52.18 mm
52.18 mm
Signal quality
78%
< Previous | Next >
13-04-2010 14:11:09 13 MB free
CAUTION!
If the actual sensor position is not entered accurately, the sound velocity can not be calculated
correctly and optimization of the sensor position will fail.
1.10
Advised sensor position
Actual sensor position
66.05 mm
66.05 mm
< Previous | Next >
13-04-2010 14:11:09 13 MB free
1.12
Volume flow
Velocity of sound
Signal quality
Optimize position
Optimize position
Optimize position
Optimize position
50 m
3
/h
1481.3 m/s
69%
< Previous | Next >
13-04-2010 14:11:09 13 MB free
INFORMATION!
The optimization loop uses the measured VoS and compares it to the entered VoS.