Technical data – KROHNE OPTISONIC 6300 P-6400 EN User Manual
Page 80
8
TECHNICAL DATA
80
OPTISONIC 6300 P
www.krohne.com
03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en
Pulse or frequency output
Pulse or frequency output
Pulse or frequency output
Pulse or frequency output
Isolation
The output is galvanically isolated from the other circuits.
Output data
For pulse counting and/or analog output: Volume flow, mass flow,
thermal energy (requires input of temperature (2x)
As analog output: Flow speed, speed of sound, gain, signal to noise ratio,
reliability of flow measurement, quality of acoustic signal
Function
Can be set as a pulse output or frequency output
Settings
For Q = 100%: 0.01...10000 pulses per second or pulses per unit volume
Pulse width: setting automatic, symmetric or fixed (0.05...2000 ms)
Operating data
Active
U
nom
= 15 VDC
ffff
max
max
max
max
≤ 100 Hz:
100 Hz:
100 Hz:
100 Hz:
I ≤ 20 mA
open:
I ≤ 0.05 mA
closed:
U
0,nom
= 15 V at I = 20 mA
100 Hz < f
100 Hz < f
100 Hz < f
100 Hz < f
max
max
max
max
≤ 10 kHz:
10 kHz:
10 kHz:
10 kHz:
I ≤ 20 mA
open:
I ≤ 0.05 mA
closed:
U
0,nom
= 13.5 V at I = 1 mA
U
0,nom
= 12.5 V at I = 10 mA
U
0,nom
= 9 V at I = 20 mA
Passive
U
ext
≤ 32 VDC
ffff
max
max
max
max
≤ 100 Hz:
100 Hz:
100 Hz:
100 Hz:
I ≤ 100 mA
open:
I ≤ 0.05 mA at U
ext
= 32 VDC
closed:
U
0, max
= 0.2 V at I ≤ 10 mA
U
0, max
= 2 V at I ≤ 100 mA
100 Hz < f
100 Hz < f
100 Hz < f
100 Hz < f
max
max
max
max
≤ 10 kHz:
10 kHz:
10 kHz:
10 kHz:
I ≤ 20 mA
open:
I ≤ 0.05 mA at U
ext
= 32 VDC
closed:
U
0, max
= 1.5 V at I ≤ 1 mA
U
0, max
= 2.5 V at I ≤ 10 mA
U