Installation for flow measurement, 5 connect the sensor cable, Optisonic 6300 p – KROHNE OPTISONIC 6300 P-6400 EN User Manual

Page 30: Signal strength qualification

background image

3

INSTALLATION FOR FLOW MEASUREMENT

30

OPTISONIC 6300 P

www.krohne.com

03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en

3.4.5 Connect the sensor cable

Figure 3-16: Connecting signal cables

1 Connector for "UP" transducer (blue)
2 Connector for "DOWN" transducer (green)

1

2

3

4

5

6

7

8

9

+/-

0

EXP

ABC

DEF

GHI

JKL

MNO

PQRS

TUV

WXYZ

A/a

-

I/O

DISPLAY

LOGGER

MENU

1

1

2

2

1

2

INFORMATION!
After mounting the transducers in the proper position, it takes up to 60 seconds to find a signal.

INFORMATION!
The signal strength is given and can be qualified as shown:

Signal strength

Qualification

> 75%

High signal

50...75%

Fairly high signal

10...50%

Low signal

< 10%

Bad or no signal

Check settings in menu, change transducer distance

until there is at least a low signal.