GETTING STARTED

9

PCS-1000 Overview

The PCS-1000 uses five high-precision shunt resistors as the basis
for accurate current and voltage measurements. The 5 shunt ranges
are 0.001Ω, 0.01Ω, 0.1Ω, 1Ω, 10Ω with a current measurement range
of 300A, 30A, 3A, 300mA and 30mA, respectively.

Main Features

Performance

Wide DC/AC voltage range
(200mV ~ 600VAC/1000VDC)

Wide AC/DC current range (30mA ~ 300A)

Low drift at all ranges

Low temperature coefficients

Features

Shunts: 0.001Ω, 0.01Ω, 0.1Ω, 1Ω, 10Ω

Current Meter (6 1/2 digits current meter)

Voltage Meter (6 1/2 digits voltage meter)

Current Monitor

Voltage and current can be measured at the
same time.

Interface

USB

GPIB for SCPI commands